Abstract

Single needle dialysis is used as a method of vascular access in many European dialysis centres. However, recirculation has been incriminated as a disadvantage of this procedure. According to in vitro studies, recirculation should be limited to a minimum when using double lumen needles. The present study makes an in vitro comparison between single and double lumen needles in unipuncture (single needle) dialysis of the type double headpump (Bellco, BL 760) and studies the role of recirculation in dialysis performance. Although recirculation is significantly reduced with the double lumen needle, compared to the single lumen needle, the clearance of small molecules is not significantly different for both needle types.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.