Abstract
Abstract The application of bioglass coatings on metallic implants provides link between bone and materials and prevents corrosion of metallic implants in body fluid. Therefore, in this research, 45S5 bioglass-silica coatings on 316L stainless steel were prepared by the sol–gel method and were characterized by different techniques. According to X-ray diffraction (XRD) results, by sintering 45S5 bioglass at 600 °C for 5 h, coatings containing both amorphous phase and Na2Ca2Si3O9 crystalline phase were obtained. Scanning electron microscopy (SEM) results showed that coatings prepared via appropriate sol aging and substrate preparations are crack-free. Potentiodynamic polarization tests in simulated body fluid (SBF) showed that coatings improve corrosion resistance of substrates, significantly. Further, bioactivity was considered during 30 days of immersion in SBF at 37 °C by atomic absorption spectroscopy (AAS), XRD, SEM and potentiodynamic polarization tests that demonstrated the growth of amorphous apatite, hydroxyapatite (HA) and CaSiO3 film on coatings.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.