Abstract

To clarify the dosimetric characteristics of a real-time variable shape rubber-containing tungsten (STR) bolus in a clinical plan and investigate the efficacy of the STR bolus in photon radiotherapy for keloids and other superficial tumors.A 5 mm gel bolus or 1 mm STR bolus was placed on a solid water phantom. Tangential irradiation was performed using a TomoTherapy Radixact-X9 and 6 MV X-ray flattening-filter-free beam, and the surface dose was measured with radiochromic film. Clinical-like plans (TomoDirect; TD and TomoHelical; TH) were applied with the same geometry and the dose distributions were measured.The increase in surface dose by the build-up effect and backscatter was 37.7% and 8.0% for the gel bolus, and 40.5% and 26.4% for the STR bolus, respectively. In the TD and TH plans, the increase in surface dose was 27.4% and 48.3% for the gel bolus, and 39.0% and 63.2% for the STR bolus. Similary, changes in the sagittal plane dose were - 3.9% and 6.1% for the gel bolus, and - 6.3% and 6.9% for the STR bolus.The STR bolus effectively increased the surface dose by the build-up effect and backscatter in photon radiotherapy for keloids and other superficial tumors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.