Abstract

Based on double-dosemeter readings, a conservative effective dose (E) estimation algorithm for lead apron workers in interventional radiology is proposed. Typical radiation conditions for various exposure geometries were simulated using the MCNPX 2.4.0 code. The simulation model consisted of an X-ray source and image intensifier, a patient phantom and a voxelised staff member phantom with lead apron. The effective staff dose and dosemeter readings for several positions of the worker were calculated. The effective dose to a physician, positioned in close proximity to the primary beam, can be estimated within a 10% underestimation margin by E = 1.64 H(p)(10)thorax,under + 0.075 H(p)(10)neck,over. The dose to the eye lens can be estimated by a dosemeter reading at collar level (R2 = 0.98).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.