AbstractA new way of decreasing the detection limit ‐ double deposition and stripping steps was proposed to determine trace amounts of gold(III) by anodic stripping voltammetry. Two carbon composite electrodes that differed drastically in their surface areas were used for the measurements. The calibration graph was linear from 1×10−9 to 1×10−8 mol L−1 following deposition time of 300 s at the first and the second electrode. The detection limit was found to be 2.3×10−10 and 1.4×10−11 mol L−1 for deposition time 600 and 2400 s, respectively. It is the lowest detection limit obtained so far for gold(III) determination in stripping voltammetry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.