Abstract

Theoretical relations for optimum adjustments of two arrangements of a double-bent-crystal (DBC) small-angle neutron scattering (SANS) diffractometer are discussed. Results of several experiments demonstrate the applicability of DBC SANS in the investigation of some technologically important materials in the range of scattering-vector sizes 0.0002–0.02 Å−1. Formulas derived for data analysis, which describe the transformation of the SANS cross section to the measured intensity, indicate that water calibration is not required and that the indirect method of data evaluation is the most appropriate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.