Abstract

The concept of an active dosimetry system for pulsed radiation dose rate measurements is presented. Real-time distinction of pulsed and non-pulsed radiation contributions is based on the time structure of a single interaction. A fast tissue equivalent plastic scintillator is exploited to minimize the pile-up effect influence on absorbed energy measurements. Being connected to a fully digital signal processing board, the detector creates an active dosimetry system with adjustable parameters. With this system, absorbed dose rate measurements were carried out in a photon field with a time structure mimicking a radiotherapeutic beam, but also in the presence of a constant radiation field. Measurements show a linear dependence of a pulsed radiation contribution on the accelerator current in the investigated range of the total dose rate up to 8 μGy h-1. While increasing the accelerator current by 1 μA, the pulsed radiation dose rate grows by (26.2±0.9) nGy h-1 when considering pile-up events.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.