Abstract
The dosimetry parameters of the IRH10 (192)Ir high dose rate brachytherapy source were obtained from the dose calculation formalism recommended in the AAPM Task Group No. 43 report using the Monte Carlo code PENELOPE. The absorbed doses to water and air originating from the photons of the IRH10 (192)Ir brachytherapy source were calculated by the collision kerma approximation. The dose rate constant was evaluated to be (1.110 +/- 0.011) cGy/h U(-1). The dose rate per unit air kerma strength around the (192)Ir IRH10 brachytherapy source and the anisotropy function were given in tables and figures.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.