Abstract

The dosimetry and dose verification for 6-MV X-rays were performed for radiosurgery cones of 5- to 40-mm diameter. The total scatter factors decrease slowly from 0.936 (40-mm cone) to 0.893 (10-mm cone; a variation of 5%), but they fall to 0.83 (7.5-mm cone) and 0.67 (5-mm cone). The dmax increases from about 12.9 (5-mm cone) to 16.3 mm (40-mm cone). The full width half maximum (FWHMs) of the beam profiles, measured at 5 cm depth, agree with the cone diameters within 1 mm. The 10-90% beam penumbra/FWHM ratio is 0.23 +/- 0.03 (> or = 20-mm cones); for the smaller-diameter cones this ratio increases reaching 0.84 (5-mm cone). New tissue maximum ratios (TMRs) are reported for the 5-, 7.5-, 32.5-, and 37.5-mm-diameter cones. TMRs for the other diameter cones are consistent with published data. The measured doses in two verification studies using the 12 cones with diameters > or = 12.5 mm with a single 360 degrees arc agreed to 2% with the planned doses, and to about 10% for the three smaller cones. In a simulated treatment neglecting tissue heterogeneties (skull bone), the measured doses for two five arc studies (22.5-mm cone) were within 4% of the calculated dose to isocenter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.