Abstract

ObjectiveThe present study was aimed at evaluating the effective radiation dose in patients submitted to PET/CT for the diagnosis of mild cognitive impairment.Materials and MethodsTLD-100 detectors inserted into an Alderson Rando® anthropomorphic phantom were utilized to measure the absorbed dose coming from the CT imaging modality. The anthropomorphic phantoms (male and female adult versions) were submitted to the same technical protocols for patients’ images acquisition. The absorbed dose resulting from the radiopharmaceutical injection was estimated by means of the model proposed by the ICRP publication 106.ResultsThe effective dose in patients submitted to this diagnostic technique was approximately (5.34 ± 1.99) mSv.ConclusionOptimized protocols for calculation of radioactive activity injected into patients submitted to this diagnostic technique might contribute to reduce the effective radiation dose resulting from PET/CT in the diagnosis of mild cognitive impairment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.