An X-ray dosimetry intercomparison for partial body irradiations has been performed among a number of laboratories cooperating within EULEP (European Late Effects Project Group) using mailed acrylic plastic rat phantoms. The main objective of this study was to assess the adequacy of exposure arrangements and dosimetry procedures applied to rat partial body irradiations at the participating institutes. The dose and dose distribution in lung and liver substitute material have been measured with dual thermoluminescent (TL) dosemeters LiF and CaF2:Mn. The doses from scattered radiation in the shielded parts of the phantom have also been assessed. The results of the dose intercomparison show that in the first instance six out of the eight participating institutes obtained dose values which differ by more that 5% from the standard dose. Inconsistencies are observed between the lung and liver doses administered at the same institute. This project has indicated the need for repeated intercomparisons and for improvements in dosimetry and irradiation procedures for partial body irradiations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.