Abstract

Purpose : To obtain more accurate information on the dose distribution in rhesus monkeys for total body irradiation with orthovoltage X-rays. Materials and methods : Dose measurements were performed with an ionization chamber inside homogeneous cylindrical and rectangular phantoms of various dimensions and in phantoms containing lung-equivalent material. The irradiations were carried out with reference to a monitor ionization chamber placed alongside the phantom or the irradiation cage. Results : Correction factors for mass and lung dose relative to the average dose in a homogeneous reference phantom, showed linear relationships with the effective diameter of the monkey. The lung dose correction factor relative to the homogeneous phantom was about 1.12 for a 3.5kg monkey. The stated values for the average absorbed dose in the animal of standard weight should be multiplied by a factor of 0 93 for experiments performed before 1983. All publications on total body irradiations of monkeys at TNO after 1983 contain the corrected dose values. Conclusion : Dose distributions are reported for phantoms of different diameters and of cylindrical or rectangular shape. The new dosimetry has also resulted in a revised statement of the LD 50 for the occurrence of bone marrow syndrome after X-irradiation; 4.9 Gy instead of 5.3 Gy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.