Abstract

A microbeam array was formed with a multi-slit collimator (MSC) for research on radiation therapy (MRT). Kodak EDR2 film was used to measure the dose distribution of the microbeam array. The calibration curve of optical density of the film with respect to a dose was established using a standard Farmer chamber and 60Co gamma-ray source. The peak dose of 3.6 Gy/s at the maximum was derived from the film dosimetry using the calibration curve. The uncertainty was estimated to be 5% which was mainly attributed to the uncertainty of the calibration. It was found that the ionization chamber used for monitoring the dose during the MRT experiments gave lower dose by about 30% than the dose derived from the film dosimetry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.