Abstract

A three-dimensional electron beam dose calculational algorithm implemented for use in a three-dimensional treatment planning system is described. The 3-D electron beam calculations have been in use clinically for more than seven years. The algorithm uses a pencil beam model based on small angle multiple Coulomb scattering theory. Our implementation allows volumetric CT-based inhomogeneity corrections and provides for irregular field shapes (fields shaped with cerrobend cutouts) and the use of bolus. As part of NCI-funded work evaluating the state of the art in electron beam treatment planning, extensive algorithm verification was undertaken and results of these tests for our implementation are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.