Abstract

To validate the Acuros® XB (AXB) algorithm in Eclipse treatment planning system (TPS) for RapidArc™ (RA) technique following the software upgrades. A Clinac-iX (2300CD) linear accelerator and Eclipse TPS (Varian Medical System, Inc., Palo Alto, USA) was used for commissioning of AXB algorithm using a 6 megavolts photon beam. Percentage depth dose (PDD) and profiles for field size 2 cm × 2 cm, 4 cm × 4 cm, 6 cm × 6 cm, 10 cm × 10 cm, 20 cm × 20 cm, 30 cm × 30 cm to 40 cm × 40 cm were taken. AXB calculated PDDs and profiles were evaluated against the measured and analytical anisotropic algorithm (AAA)-calculated PDDs and profiles. Test sites recommended by American Association of Physicists in Medicine task group (AAPM TG)-119 recommendation were used for RA planning and delivery verification using AXB algorithm. AXB shows excellent agreement with measurements and AAA calculated data. The CLs were consistent with the baseline values published by TG-119. AXB algorithm has the potential to perform photon dose calculation with comparable fast calculation speed without negotiating the accuracy. AAPM TG-119 was successfully implemented to access the proper configuration of AXB algorithm following the TPS upgrade.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.