Abstract

PurposeMagnetic resonance image (MRI) guided radiotherapy enables gating directly on the target position. We present an evaluation of an MRI‐guided radiotherapy system's gating performance using an MRI‐compatible respiratory motion phantom and radiochromic film. Our evaluation is geared toward validation of our institution's clinical gating protocol which involves planning to a target volume formed by expanding 5 mm about the gross tumor volume (GTV) and gating based on a 3 mm window about the GTV.MethodsThe motion phantom consisted of a target rod containing high‐contrast target inserts which moved in the superior‐inferior direction inside a body structure containing background contrast material. The target rod was equipped with a radiochromic film insert. Treatment plans were generated for a 3 cm diameter spherical planning target volume, and delivered to the phantom at rest and in motion with and without gating. Both sinusoidal trajectories and tumor trajectories measured during MRI‐guided treatments were used. Similarity of the gated dose distribution to the planned, motion‐frozen, distribution was quantified using the gamma technique.ResultsWithout gating, gamma pass rates using 4%/3 mm criteria were 22–59% depending on motion trajectory. Using our clinical standard of repeated breath holds and a gating window of 3 mm with 10% target allowed outside the gating boundary, the gamma pass rate was 97.8% with 3%/3 mm gamma criteria. Using a 3 mm window and 10% allowed excursion, all of the patient tumor motion trajectories at actual speed resulting in at least 95% gamma pass rate at 4%/3 mm.ConclusionsOur results suggest that the device can be used to compensate respiratory motion using a 3 mm gating margin and 10% allowed excursion results in conjunction with repeated breath holds. Full clinical validation requires a comprehensive evaluation of tracking performance in actual patient images, outside the scope of this study.

Highlights

  • Radiation therapy provides an effective way to combat numerous types of cancer, improving both a patient’s quality and quantity of life

  • We present a dosimetric evaluation of the MRIdian’s respiratory gating performance using an MRIcompatible motion phantom and radiochromic film

  • Considering all trajectories, increased accuracy is observed for 3 mm versus 5 mm gating margin (P < 0.004) and for 5% allowed excursion versus 10% allowed excursion (P = 0.005)

Read more

Summary

Introduction

Radiation therapy provides an effective way to combat numerous types of cancer, improving both a patient’s quality and quantity of life. Our methodology is to perform static, gated moving, and ungated moving deliveries of the same plan and evaluate the gating performance in terms of its ability to “freeze” motion, i.e., quantitate the difference between each case and the static treatment plan using the gamma[16] metric. This methodology has been frequently used to evaluate gating systems[17,18,19,20,21,22,23]. The purpose of this study was to establish a technical performance baseline of the respiratory gating which can inform clinical gating protocols as well as be used as a machine performance benchmark by other institutions

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.