Abstract
Tumor motion due to patient's respiratory is a significant problem in radiotherapy treatment of lung cancer. The purpose of this project is to study the interplay effect through dosimetry verification between the calculated and delivered dose, as well as the dosimetric impact of leaf interplay with breathing-induced tumor motion in IMRT and VMAT treatment. In this study, a dynamic thorax phantom was designed and constructed for dosimetry measurement. The phantom had a linear sinusoidal tumor motion toward superior-inferior direction with variation of amplitudes and periods. TLD-100 LiF:Mg,Ti and Gafchromic EBT2 film were used to measure dose in the midpoint target and the spinal cord. The IMRT and VMAT treatment had prescription dose of 200 cGy per fraction. The dosimetric impact due to interplay effect during IMRT and VMAT treatment were resulted in the range of 0.5% to -6.6% and 0.9% to -5.3% of target dose reduction, respectively. Meanwhile, mean dose deviation of spinal cord in IMRT and VMAT treatment were around 1.0% to -6.9% and 0.9% to -6.3%, respectively. The results showed that if respiratory management technique were not implemented, the presence of lung tumor motion during dose delivery in IMRT and VMAT treatment causes dose discrepancies inside tumor volume.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.