Abstract

A new brachytherapy (125)I source has been investigated at Iranian Agricultural, Medical and Industrial Research School. Dosimetric characteristics [dose-rate constant Λ, radial dose function g(l)(r) and anisotropy function F(r,)] of IRA-(125)I were theoretically determined in terms of the updated AAPM task group 43 (TG-43U1) recommendations. Versions 5 and 4C of the Monte Carlo radiation transport code were used to calculate the dosimetry parameters around the source. The Monte Carlo calculated dose-rate constant of the (125)I source in water was found to be 92×10(-4) Gy h(-1) U(-1) with an approximate uncertainty of ±3 %. Brachytherapy seed model, 6711-(125)I, carrying (125)I radionuclides, was modelled and benchmarked against previously published values. Finally, the calculated results were compared with the published results of those of other source manufacturers.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call