Abstract

There is limited study comparing dosimetry parameters in detail. In regard to prostate cancer, there are four different techniques, namely three-dimensional conformal radiotherapy (3D-CRT), intensity-modulated radiotherapy-step and shoot (IMRT-SS), IMRT-helical tomotherapy (HT), and volumetric-modulated arc therapy (VMAT). Experimental study with intervention on ten prostate cancer patients' computed tomography planning data. 78 Gy dose in 39 fractions was given for planning target volume.Experimental study with intervention on ten prostate cancer patients' computed tomography planning data. 78 Gy dose in 39 fractions was given for planning target volume. The mean V75 Gy rectum and bladder between 3D-CRT and the other three abovementioned techniques all showed significant results (P < 0.05). V5 Gy remaining volume at risk (RVR) between 3D-CRT versus VMAT and HT, IMRT-SS versus HT, and VMAT versus HT is statistically significant (P < 0.0001). The longest radiation time was done with HT (mean 4.70 ± 0.84 min). V75 Gy rectum bladder between 3D-CRT techniques differ significantly compared to the three other techniques and may not be suitable to the implementation of escalation doses. The HT technique produced the highest V5 Gy RVR and needed the highest monitor unit amount and the longest radiation duration. The VMAT technique was considered capable of realizing dose escalation in prostate cancer radiotherapy by minimizing toxicity in the rectum and bladder with the shortest radiation duration.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call