Abstract
We have evaluated the performance of stereo and orthogonal computerized three-dimensional implant reconstruction algorithms. Sets of orthogonal and stereo radiographs of a simulated interstitial seed implant were taken. Computer reconstructed coordinates of each source were compared with their measured positions in order to assess the geometric accuracy of the two methods. We find that the effect of random measurement errors on reconstructed coordinates from stereo films is 3–5 times that of the orthogonal method. Although relataive individual seed positions may be in error by as much as 1 cm, these errors do not significantly change the shape or area of isodose curves surrounding the implant periphery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Radiation Oncology, Biology, Physics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.