Abstract

Previous studies measured the effects of 222Rn on various thermoluminescent dosemeters (TLDs). This study quantified the effects of 222Rn on LiF : Ti,Mg chips encapsulated in plastic extremity rings. For 28 d, one batch of TLDs was left in a chamber with high radon levels, and another batch in a control chamber with normal background radon levels. A few TLDs in each batch were removed from the rings for direct exposure to the ambient air in each chamber. Passive continuous radon monitors (CRMs) recorded the 222Rn levels. TLDs were processed using a third-party dosimetry company, CRM data were analysed, and the relationship between integrated 222Rn concentration and TLD response was determined. The batch of TLDs in the experimental chamber showed a weak response to 222Rn gas, which was in the order of 0.5 nSv Bq−1 m3 d−1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.