Abstract

Diagnosis by X-ray image are procedures widely used to improve the diagnosis or to follow the evolution of a medical procedure, also are used to support the cancer treatment with linear accelerators. The procedure involves the X-ray unit and the detection system, while the X-ray beam is projected onto the patient; along this procedure X-rays are scattered out from the patient body and X-rays leak-out from the X-ray unit. Both, the scattered and the leaking out radiation, are important during the evaluation or the design of the shielding barriers. X-ray tube features like voltage, electric current, target and filter are related to the dose due to the scattered and leakage radiation. Also, the dose due to the scattered radiation depends upon the scattering angle; while the dose due to the leaking out radiation, depend upon the X-ray tube housing. The objective of this work was to estimate, using Monte Carlo methods, the X-ray spectra and doses (air Kerma and Ambient dose equivalent) produced by scattered and leaking out radiation of 70, 80 and 90 kV X-ray units.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.