Abstract
Research conducted to-date, makes use of the IBA-Lynx scintillating screen and radiochromic film to analyze the proton field uniformity for dose-area product (DAP) determination. In this paper, the machine log file based reconstruction is proposed to calculate the field uniformity to simplify the measurement. In order to calculate the field uniformity, the dose distribution is reconstructed based on the machine log file with matRad (an open source software for analytical dose calculation in MATLAB). After acquisition of the dose distribution, the field flatness and symmetry are calculated automatically for different proton energies. A comprehensive comparison of DAP determined with Bragg peak chamber (BPC) and Markus chamber (MC) is presented. The actual delivered dose is reconstructed with the log file to analyze the lateral dose distribution of the scanned field. DAP of different energies are calculated ranging from 70.6 MeV to 235 MeV. The percentage difference is calculated, illustrating the DAP discrepancy between the MC and BPC to the mean value. The percentage difference ranges from −0.19% to 1.26%. The variation between DAP measured with the BPC and MC peaks at −2.5%. The log file based reconstruction to calculate field uniformity can be an alternative for DAP determination. The direct method using a large-area Bragg peak chamber is investigated. The two methods to determine DAP and calibrate beam monitor illustrate consistent results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.