Abstract

Improvement of the dose homogeneity in radiation treatment of the intact breast using 3D-planning and dose volume histograms. 3D-planning, including the calculation of dose volume histograms of the planning target volume, was performed on 15 patients, who underwent radiation therapy with tangential photon beams. A standard plan and 2 modified or optimized plans were evaluated. Different dosimetric parameters like maximum dose, mean dose, standard deviation and the fractional volume which receives doses from 95 to 105% of the reference dose were compared and correlated with breast size. With increasing breast size standard planning leads to increased overdosage, both in magnitude and volume. Individual optimization by modifying weights and wedges gives no improvement in dose homogeneity, whereas a photon energy of 10 MV results in a more homogeneous dose distribution. The drawback of the higher energy is the increased underdosage of the skin. Using the standard geometry of tangential fields the dose homogeneity cannot be improved significantly by 3D-planning, compared to our standard technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.