Abstract

This work presents a dosimetry system based on MOSFET sensors for real-time dose monitoring. MOS transistors were biased during irradiation, and the response of lateral, general-purpose 3N163 and CD4007 transistors were characterized with a 15-MV photon beam provided by a linear accelerator. The electronic circuitry to condition the sensor output and the measurement algorithm are described in depth. Due to the real-time measurement mode, the dosimetric parameter (the source voltage) showed drift. This drift depends on the bias voltage applied between the gate and the bulk terminals alternately during irradiation and readout. It can be minimized by applying a 1-V bias voltage for the 3N163 transistor and 0.85V for the CD4007 during readout. The CD4007 transistor showed an average sensitivity of (7.8±0.4)mV/Gy and the 3N163 an average of (26.4±0.8)mV/Gy. The low uncertainty and acceptable sensitivity yields a resolution of 0.8 and 1.5cGy for the CD4007 and 3N163 transistors respectively for unirradiated devices, which increases to 3cGy and 2cGy after 16Gy of cumulative dose, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.