Abstract

Thermoluminescent dosimeters are well established for performing calibrations in radiotherapy and for monitoring dose to personnel exposed to low linear energy transfer (LET) ionizing radiation. Patients undergoing light ion therapy and astronauts engaged in space flight are, however, exposed to radiation fields consisting of a mix of low- and high-LET charged particles. In this study, glow curves from chips were examined after exposure to various electron and ion beams. The annealing and readout procedures for these chips were optimized for these beams. After a 10 min prereadout annealing at , the optimized glow curve samples the light output between 95 and with a heating rate of . The ratio of the integral of the glow curve under peaks 4–6 to the integral under peak 3 was approximately 0.9 for electrons, 1.0 for entrance protons, 1.6 for peak protons, and 2.2 for entrance carbon, silicon, and iron ions. The integral light output per unit dose in water for the iron exposures was about half as much as for the electron exposures. The peak-area-ratio can be used to determine a dose response factor for different LET radiations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.