Abstract

Results of characterization studies on one of the first versions of the Sunna photo-fluorescent dosimeter™ have previously been reported, and the performance of the red fluorescence component described. This present paper describes dose response and post-irradiation characteristics of the green fluorescence component from the same dosimeter film ( Sunna Model γ), which is manufactured using the injection molding technique. This production method may supply batch sizes on the order of 1 million dosimeter film elements while maintaining a signal precision (1 σ) on the order of ±1% without the need to correct for variability of film thickness. The dosimeter is a 1 cm×3 cm polymeric film of 0.5-mm thickness that emits green fluorescence at intensities increasing almost linearly with dose. The data presented include dose response, post-irradiation growth, heat treatment, dosimeter aging, dose rate dependence, energy dependence, dose fractionation, variation of response within a batch, and the stability of the fluorimeter response. The results indicate that, as a routine dosimeter, the green signal provides a broad range of response at food irradiation (0.3–5 kGy), medical sterilization (5–40 kGy), and polymer cross-linking (40–250 kGy) dose levels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.