Abstract

AIM: To assess the effectiveness and economy of routinely using commercially available in-plane bismuth shielding during CT scanning of the chest and brain. METHODS AND MATERIALS: Forty patients were scanned with thermoluminescent dosemeters (TLDs) in situ to measure the radiation dose to the thyroid and eye during CT scanning of the brain and chest. Half of the patients had the “AttenuRad” shield in place during scanning. RESULTS: Use of the shielding reduced the mean dose to the eye from 6.0±0.3 mGy to 4.9 mGy±0.2 mGy and the thyroid dose from 16.4 mGy±1.2 mGy to 7.1 mGy±0.5 mGy. CONCLUSION: Use of the thyroid shield is recommended for all CT scanning of the chest. The eye shield does not produce as marked a reduction in radiation dose to the lens of the eye, when an angled gantry is used, since the eyes are not in the primary beam. Use of the eyeshield is justifiable where irradiation of the orbit is unavoidable, although whether artifacts would be a problem if the shield was used in this way was not assessed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.