Abstract
A modern programmable panoramic radiography unit was used to irradiate an anthropomorphic phantom in order to assess the variation in organ absorbed dose and calculated effective dose with program selection. The results of the measurements for five of the 12 programs available are presented, together with those obtained from an older, non-programmable unit. The results of dosimetry measurements are compared between units, between programs, and with previous data. Variations in radiation dose and their implications are discussed, and the opportunities for reductions in organ dose and effective dose of up to 85% and 50%, respectively, demonstrated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.