Abstract

To demonstrate fluoroscopy dose reduction through both simulated injections on a phantom and patient injections. Our study was IRB-approved and HIPAA-compliant. Simulation on a phantom was used to estimate effective dose, entrance dose, and organ doses for hip joint injections without and with dose minimizationtechnique (DMT). Additionally, 1,094 joint, bursae, and tendon sheath injections performed by three operators in the same fluoroscopy suite were evaluated both before and after application of DMT. Fluoroscopy time (FT), dose, and dose area product (DAP) of injections were compared using unpaired t-tests with P>0.05 considered statistically significant. For the phantom simulation comparing injections without DMT and with DMT, the total DAP was 191.7 vs 18.7 μGy·m2, and the entrance dose was 10.2 vs 3.6mGy, respectively. For both men and women, DMT reduces effective dose and organ doses. For all injections, the FT (0.7 to 0.2min), dose (5.6 to 1.9mGy), and DAP (56.9 to 19.1 μGy·m2) for all three operators decreased with DMT and remained statistically significant when stratified by the two most common injections, glenohumeral and hip joint injections (P<0.05). FT, effective dose, entrance dose, and DAP can be reduced with the use of simple easy-to-learn techniques, which will benefit both the patient and the radiologist.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call