Abstract

This work, using an adult anthropomorphic phantom, aimed to establish an optimised technique for ladies of child-bearing age undergoing antero-posterior (AP) pelvis and AP and lateral lumbar spine examinations. Phase one of the work involved introducing the following dose-reducing measures individually: increased kVp, increased focus-film distances, a carbon fibre cassette, a faster film/screen combination. The second phase established an optimised technique based on a combination of the parameters listed above. Radiation dose was measured using thermoluminescent dosimeters and image quality was evaluated using anatomical criteria. All dose-reducing methods were compared with a standard technique, currently being used in a Dublin hospital. The results demonstrated that the optimised procedure reduced effective dose by 77, 62 and 66% for AP pelvis and AP and lateral lumbar spine respectively (p < 0.05) compared with the standard technique, with no significant changes in image quality. Dose-reducing measures used in combination offer substantial potential for optimisation of radiological procedures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.