Abstract

We demonstrate a practical calibration method and its applicability for a commercially available radiophotoluminescence dosemeter (RPLD), i.e. the GD-352M (AGC Techno Glass, Shizuoka, Japan) to eye lens dose monitoring, by performing the calibration according to the ISO recommendations. The calibration was then verified through a series of experiments. For verification of the derived calibration factor (1.21 ± 0.04, k = 1) of the RPLD, we performed standard irradiations in the ISO narrow series X-ray reference fields and the simulation measurements in the actual radiation fields in a hospital. The TLD-based commercially available dosemeters, DOSIRIS™ was also put on the ISO cylinder phantom and the RANDO phantom together with the GD-352M in the verification experiments. The personal dose equivalents Hp(3) obtained from the GD-352M and those obtained from the DOSIRIS™ were in good agreement with each other. Our results demonstrate the proper calibration of a commercially available RPLD that is applicable to the additional monitoring of the lens of the eyes for medical staff.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.