Abstract
Inter- and intra-fractional prostate motion can deteriorate the dose distribution in extremely hypofractionated intensity-modulated proton therapy. We used verification CTs and prostate motion data calculated from 1024 intra-fractional prostate motion records to develop a voxel-wise based 4-dimensional method, which had a time resolution of 1s, to assess the dose impact of prostate motion. An example of 100 fractional simulations revealed that motion had minimal impact on planning dose, the accumulated dose in 95% of the scenarios fulfilled the clinical goals for target coverage (D95>37.5Gy). This method can serve as a complementary measure in clinical setting to guarantee plan quality.