Abstract

Radioprotection in dental radiology is of particular interest in the framework of the Revised Medical Exposure Directive for the great number of examinations involving the adult as well as the paediatric population (Article 9: Special Practice). The present study is intended to find the quality criteria of orthodontic imaging and for evaluating the dose absorbed within the dental and maxillary volume in connection with radiological examinations performed with either spiral CT, dental panoramic tomography or teleradiography. The X ray dose to organs sited in the body, neck, ocular and intracranial area was measured using lithium fluoride dosemeters, positioned in a Rando phantom. Quality criteria have been established by an expert radiologist considering the diagnostic information obtained in the images. The dosimetric data obtained were comparable with other authors', although with some differences due to technical characteristics. These result data are useful for choosing the patient's diagnostic path, considering the radiobiological risk associated with increasing orthodontic imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.