Abstract

Calibration factors for commercial ionization chambers (i.e. dose calibrators) were determined for a solution of 123I; the activity was based on the 1976 NBS standard. A link between the NIST standard and the International Reference System (SIR) was established. The two major U.S. dose calibrator manufacturers recommend oppositely biased calibration factors, giving a spread of 11.3% in measured activities. With modern quantitative imaging techniques capable of ≤10% accuracy, this bias for a SPECT nuclide is highly significant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.