Abstract
In this study, diagnostic reference levels (DRLs) were suggested and patient doses were analysed through the dose-area product value in dental radiography. In intraoral radiography, at three sites, i.e. molar, premolar and incisor on the maxilla and acquired third quartile values: 55.5, 46 and 36.5 mGy cm(2), respectively, were measured. In panoramic, cephalometric and cone beam computed tomography, the values were 120.3, 146 and 3203 mGy cm(2) (16 × 18 cm), respectively. It has been shown that, in intraoral radiography, the patient dose changes proportionally to the value of mA s, but the change in extraoral radiography in response to mA s could not be confirmed. The authors could confirm, however, the difference in dose according to the manufacturer in all dental radiography examinations, except for panoramic radiography. Depending on the size of hospital, there were some differences in patient dose in intraoral radiography, but no difference in patient dose in extraoral radiography.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.