Abstract

With single-slice helical CT, an increased pitch can decrease the radiation dose to the patient if all other parameters are constant. The purpose of this study was to determine whether the same relationship holds for a particular multislice helical CT system (Somatom Plus 4 VZ multislice helical CT scanner, version A11A) in our department. The measured radiation dose to the phantom was identical for all pitch selections on the multislice helical CT system we tested. This unexpected result was because of an automatic proportionate increase in the tube current when the pitch selection was increased. Radiologists and physicists should exercise caution when extrapolating dose reduction strategies from single-slice to multislice helical CT systems, and they must acquire a detailed understanding of the multislice helical CT scanner of their chosen manufacturer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.