Abstract
An X-ray fluorescence method is described for the determination of plutonium in Pu-Al alloys. In order to obtain satisfactory reproducibility, the alloy is first dissolved in a nitric acid-mercury (II) nitrate mixture. The Lα emission of plutonium is measured, to eliminate matrix effects, an internal standard with a wavelength near that of the Pu emission is used. Equipment and sample preparation are described. Extension of the method to other determinations is discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.