Accurate assessment of atrioventricular conduction time (AVCT) is crucial for early detection of evolving fetal heart block. Simultaneous pulsed wave Doppler (PD) interrogation of left ventricular inflow and outflow is mainly used to study fetal AVCT. Limitations of this modality include its dependency on loading conditions and merging early and late diastolic inflow waves at faster heart rate (HR). Sequential analysis of atrioventricular myocardial motion by Doppler tissue imaging (DTI) might be more useful in this regard. In 15 open-chest pigs, AVCT was measured by PD, DTI, and electrocardiogram at baseline HR and during incremental atrial pacing up to 200 beats/min. Electromechanical delay and pre-ejection period were assessed at baseline and maximal HR. DTI-derived AVCT correlated better with PR intervals and allowed measurements at faster HRs than did PD (P < .05). Pre-ejection period prolonged with faster HR (P < .001), unlike electromechanical delay. In conclusion, DTI allows more accurate measurement of AVCT over a wider HR range than does PD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.