Abstract

Serial Doppler echocardiographic examinations were carried out in random order on six boxer dogs, on 3 separate days, by two experienced Doppler echocardiographers, to assess measurement variability and reproducibility of 65 parameters. Large numbers of parameters exhibited significant differences for each of the categories of intraobserver, interobserver, interday and interoperator. The coefficients of variation for all parameters measured ranged from 5.03 to 46.43%, but most were less than 20%. In general, least variation was found for the intraobserver category, and the best reproducibility for M-mode and left ventricular volumetric data. The worst reproducibility was found for tricuspid inflow and pulmonary venous flow measurements. The results of this study suggest differences greater than 20% for serial scans must be achieved to document genuine change, although the specific data should be consulted. Furthermore, variability and reproducibility are improved if a single experienced operator/observer acquires and measures serial scans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.