Abstract

Four episodes of St. Jude Medical prosthesis leaflet thrombosis were serially evaluated by two-dimensional and Doppler echocardiography during treatment with streptokinase. Three patients aged 4, 11, and 24 years with congenital heart disease had St. Jude Medical valves in pulmonary positions (two cases) for tetralogy of Fallot and in the tricuspid position (one case). The duration of thrombosis was not known in any patient. Leaflet immobility and its resolution were demonstrated by echocardiography and were confirmed fluoroscopically. Continuous wave Doppler echocardiography showed abnormal stenotic gradients in thrombosed valves that were reduced after thrombolysis. These studies demonstrate the utility of two-dimensional and Doppler echocardiography in serial evaluation of prosthetic pulmonary and tricuspid valve thrombosis during thrombolysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.