Abstract

The resting haemodynamics of five types of small (19 mm) aortic valve prosthesis (2 bileaflet, 3 pericardial) were evaluated with Doppler echocardiography in 43 patients. Two received St Jude Medical and six CarboMedics bileaflet valves and 35 were given bioprostheses--16 Ionescu-Shiley, four Mitroflow and 15 Labcor-Santiago. No significant differences in peak or mean transvalvular pressure drop or in effective valve area were found between the bileaflet and the pericardial valves or among the three types of bioprosthesis. All but one of the bileaflet prostheses showed a characteristic regurgitation pattern, with two lateral and one central jet, and 16 (46%) of the bioprostheses showed central regurgitation, but in no case were these jets haemodynamically significant. Thus the 19 mm bileaflet and the studied pericardial prostheses all have satisfactory resting haemodynamics, and all are suitable for implanting in small aortic roots.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.