Abstract

A new liquid chromatography–electrospray ionization-mass spectrometry (LC–ESI–MS) ( n) ion trap method for the determination of corticosteroids in urine has been developed and validated. Some anabolic agents, such as epitrenbolone, trenbolone, 2-hydroxymethylformebolone, tetrahydrogestrinone, gestrinone and formoterol were included in the LC–ESI–MS method. Matrix interference, specificity, identification capability, carry over and robustness were estimated as validation parameters. Recoveries ranged from 74 to 113% at the minimum required performance limit (MRPL), which is 30 ng mL −1 for corticosteroids and 10 ng mL −1 for anabolic agents. Methods for the confirmation and quantification of norpseudoephedrine, ephedrine, methylephedrine, salbutamol, morphine and morphine glucuronide were also developed and validated and in order to minimize analysis time, direct urine injection was used. These methods proved to be specific, accurate and precise across a calibration range for each substance since matrix interference, specificity, carry over, within and between run precision, limit of detection, limit of quantification, intermediate precision and uncertainty were estimated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.