Vacuum centrifugal (SpeedVac) concentration is commonly applied to nucleic acids extracted from formalin-fixed paraffin-embedded (FFPE) sections, but with an unknown impact. We concentrated DNA and RNA from FFPE biospecimens using different time/temperature SpeedVac combinations of up to 30 min concentration at 45 °C, then used spectrophotometry, spectrofluorometry, RIN, PERM, DV200, qRT-PCR, DIN and the Illumina FFPE QC Assay to assess the changes in quantity, purity and integrity induced by the concentration process. We found the effects of SpeedVac concentration to be inconsequential, but an aliquot of elution buffer should be concentrated for use as the blank in spectrophotometry assays.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.