Abstract

Knowledge that enables the accurate simulation of drug deposition in the human upper airway is necessary to develop robust platforms for efficient drug delivery by inhalation devices. The human upper airway is deformable during inhalation but how it could affect the deposition of inhaled drugs is unknown. We aimed to determine whether pharyngeal deformation at the soft palate level would have any significant effects on throat deposition, in vitro lung dose and fine particle fraction. In this study, dry mannitol powders were delivered to the next-generation cascade impactor (NGI) through the United States Pharmacopeia (USP) throat, and a realistic upper airway cast (RUPAC) at flow rates of 40, 60 and 80 L min-1. Deformation of the upper airway at 25%, 50%, and 75% in the lateral and antero-posterior directions were experimentally simulated in the RUPAC. Throat deposition (p = 0.04) is significantly affected when the upper airway deforms laterally but not antero-posteriorly.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.