Abstract

The aim of the study was to evaluate the incidence of peripheral inserted central catheter (PICC) tip malposition when the catheter is inserted under real-time ultrasound (RTUS) guidance when compared with conventional landmark (CL) technique in neonates. Additional objectives were to evaluate the PICC longevity and central line associated blood stream infections (CLABSI). In this randomised controlled trial, neonates were randomised to 'RTUS' (n = 40) or 'CL' (n = 40) groups. PICC tip was placed under ultrasound guidance in lower third of superior vena cava in the RTUS group. In 'CL' group, PICC was inserted as calculated by anatomical landmarks. The birth weight (1286 (926, 1662) vs. 1061 (889, 1636) g) and gestation (31.12 (3.1) vs. 31.4 (3.6) wks) were comparable among the groups. RTUS guidance during PICC insertion reduced incidence of tip malposition by 52% (67.5 vs. 32.5%; RR: 0.48; 95% CI: 0.29-0.79). The longevity of PICC and episodes of CLABSI were however similar in the two groups. Real-time ultrasound guidance during PICC placement reduces the incidence of tip malposition.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.