Abstract
Transfusions of red blood cells (RBCs) are frequently administered for a variety of clinical conditions. The use of an infusion pump has become common practice in nursing. Lack of evidence regarding hemolysis associated with transfusing older RBCs using an infusion pump led one 361-bed acute care suburban medical center to maintain its nursing policy to infuse all blood products by gravity. A combined simulated and clinical study was conducted with 25 units of RBCs. No significant hemolysis was demonstrated as measured by hemolysis index and serum potassium levels. As a result of this study, the nursing policy at the medical center was changed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of infusion nursing : the official publication of the Infusion Nurses Society
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.