Abstract

The impact of spectral filtration in digital ('cine') acquisition was investigated using a flat panel cardiac interventional X-ray imaging system. A 0.1-mm copper (Cu) and 1.0-mm aluminium (Al) filter added to the standard acquisition mode created the filtered mode for comparison. Image sequences of 35 patients were acquired, a double-blind subjective image quality assessment was completed and dose-area product (DAP) rates were calculated. Entrance surface dose (ESD) and effective dose (E) rates were determined for 20- and 30-cm phantoms. Phantom ESD fell by 28 and 41 % and E by 1 and 0.7 %, for the 20- and 30-cm phantoms, respectively, when using the filtration. Patient DAP rates fell by 43 % with no statistically significant difference in clinical image quality. Adding 0.1-mm Cu and 1.0-mm Al filtration in acquisition substantially reduces patient ESD and DAP, with no significant change in E or clinical image quality.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.