Abstract

The photon beam quality factor Q of 6- and 15-MV photon beams has been measured using different types of thimble ionization chambers with various wall, electrode and insulator materials. Some types of chambers seem to overestimate the corresponding Q-value by 0.4% (6 MV) and 0.7% (15 MV), respectively. The effect on the stopping power ratio water to air and on the pertubation correction will be discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.