Abstract

BackgroundThe posterior plating technique could be used as a clinical alternative to parallel plating for treating comminuted distal humerus fractures (DHFs) successfully with good clinical results. However, the biomechanical characteristics for posterior fixation are still unclear. The purpose of this study is to evaluate the biomechanical properties of the posterior fixation and to make comparisons between the parallel and the posterior fixation systems.Materials and methodsWe performed a cadaveric biomechanical testing with two posterior plating systems (a posterior two plating and a single posterior pre-contoured Y plating system) and one parallel two plating system to treat AO/OTA type-C2.3 DHFs. Among three groups, we compared construct stiffness, failure strength, and intercondylar width changes after 5000-cycle fatigue loading and failure loads and failure modes after destructive tests in both the axial compression and (sagittal) posterior bending directions. The correlations between construct failure loads and bone marrow density (BMD) were also compared.ResultsIn axial direction, there were no significant differences in the stiffness and failure load between the posterior and the parallel constructs. However, in sagittal direction, the two-plate groups (posterior two plating and parallel plating group) had significant higher stiffness and failure loads than the one-plate group (single posterior Y plating). There was no fixation failure after 5000-cyclic loading in both directions for all groups. Positive correlation was noted between BMD and failure loads on parallel fixation.ConclusionsWe found that when using two plates for treating comminuted DHFs, there were no significant differences in terms of most biomechanical measurements between posterior and parallel fixation. However, the single pre-contoured posterior Y plate construct was biomechanically weaker in the sagittal plane than the parallel and the posterior two-plate constructs, although there was no fixation failure after the fatigue test for all groups regardless of the fixation methods.Level of evidenceBiomechanical study

Highlights

  • Comminuted distal humerus fractures (DHFs) comprise one of the most challenging treatment situations in elbow injuries [1]

  • Positive correlation was noted between bone marrow density (BMD) and failure loads on parallel fixation

  • We found that when using two plates for treating comminuted DHFs, there were no significant differences in terms of most biomechanical measurements between posterior and parallel fixation

Read more

Summary

Introduction

Comminuted distal humerus fractures (DHFs) comprise one of the most challenging treatment situations in elbow injuries [1]. The surgical treatment principle for a distal humerus fracture is bicolumnar fixation with plate and screws [2, 3]. Current popular fixation methods for comminuted DHFs are double plating in either parallel or orthogonal direction, which have been studied widely in both clinical and biomechanical aspects [7,8,9]. Orthogonal plating is reported to have inadequate fixation in cases with osteoporotic and comminuted fractures, which may lead to fixation failure or non-unions [10]. The posterior plating technique could be used as a clinical alternative to parallel plating for treating comminuted distal humerus fractures (DHFs) successfully with good clinical results. The purpose of this study is to evaluate the biomechanical properties of the posterior fixation and to make comparisons between the parallel and the posterior fixation systems

Objectives
Methods
Results
Discussion
Conclusion